ISO 9239 Flooring Radiant Panel Tester

2022-10-12

ISO 9239 详情.png

  Introduction:   

Critical Radiant Flux is the main parameters of the fire risk for flooring materials, it reflects the burning and spreading ability of flooring material specimen under external heat radiation conditions. It is an important index of combustion classification for building products.

According to ISO9239-1,2 and GB/T 11785-2005 requirements, Flooring Radiant Panel Test Apparatus is measuring the critical radiant of combustion floor coverings, through the radiation environment of the test chamber. It is can also be used to measure the critical radiation of the cellulose insulation floor material.

The specimen is located at the down horizon of 30 degrees sloped gas radiation panel per horizon surface. Pilot flame as ignition resource put a part edge of the specimen on the high-temperature side. Flooring Radiant Panel Test Apparatus record ignition time by spark, moving time which flames horizontal pass by a specimen of length.



   Application:   

1.Building Materials

2.Rail

3.Automotive Interior

4.Other



   Standard:   

• ISO 9239-1

- Flooring fire reaction test - Part 1: Combustion performance of flooring materials tested by radiant heat;

• ISO 9239-2

- Flooring fire reaction test - Part 2: Testing flame spread status under the heat flux is 25kW/m2;

 Standard test procedure:

1、Set the air flow in the exhaust stack. Remove the dummy specimen and close the door. lgnite the panel and allow the test apparatus to heat for at least one hour until the chamber temperature has stabillized.

2、 Measure the black body temperature of the radiant panel. The black body temperature shall be within土5 °C of the black body temperature recorded during the calibration. The chamber temperature shall be within 土10 °C of the chamber temperature recorded during the calibration. If the black body or chamber temperature differs by more than the given limits, adjust the gas/air input to the radiant panel. Allow the test apparatus unit to stabilise for at least 15 min before the temperatures are measured again. When the temperatures are within the limits given, the test apparatus is ready for use. If required, adjust the smoke measuring system so that its output value is equal to 100 %. Ensure that the measuring system has stabilized before starting the tests. If not, adjust it further. Check the purging air to both the lamp and the detector system and adjust if necessary.

3、 Insert the test specimen, including any underlay(s) and substrate, into the specimen holder. Place the steel bar clamps across the back of the assembly and tighten the nuts firmly or apply other fixing means. Raise the pile of textile flooring, if applicable, using a vacuum cleaner and mount the test specimen and its holder on to the sliding platform. lgnite the pilot burner, keeping it at least 50 mm away from the intended zero point of the test specimen. Move the sliding platform into the chamber and immediately close the door. This is the start of the test. Start the timing and recording devices.



• ASTM E 648

- Critical Radiant Flux of Floor standard test - covering system uses radiant heat energy;

1、Invert the sample holder on a workbench and insert the flooring system specimen. The specimen shall be backed by a nominal 13-mm thick 0.58 g/cm3 inorganic millboard. Place the steel bar clamps across the back of the specimen/millboard assembly and tighten the nuts firmly. Return the sample holder to its upright position, clean the test surface with a vacuum, and mount it on the specimen platform. Carpet specimens shall be brushed to raise the pile to its normal position.

2、 Position the pilot burner at least 50 mm away from the specimen' s test position in the chamber.

3、 Ignite the pilot burner.

4、Move the specimen into the chamber, and, within 3 s, close the door, and start the timer. This is time zero (see 3.2.8).

5、①Allow the specimen to heat for 5 min in this position.

② If the specimen ignites before the end of the 5 min heating period, record the time of occurrence and continue with the full heating period.

6、 After 5 min, bring the pilot burner flame into contact with the specimen at the 0 mm mark while the pilot burner remains on.



• ASTM E 970

- Using a Radiant Heat Energy Source, test insulation floor insulation standard test for Critical Radiant Flux Using a Radiant Heat Energy

• NFPA 253

- Standard Test Method for using radiant heat energy, test plates overlay system Critical Radiant Flux;

• GB/T 11785-2005



   Main Features:   

1.  All structural components are made of stainless steel, beautiful appearance and corrosion resistance.

2.  The test chamber is made of the panel (calcium + silicate ) as 13±1 mm of thickness and 650 kg/m3 of nominal density.

3.  Air pass dimension between the test chamber and test specimen holder is produced by (0.23±0.03) m3 at both beside of specimen fairly.

4.  The calorimeter is imported from the United States, range 50kW/m2.

5.  The radiant chamber had a stop of 30±1degree per horizon panel and can be resisted up to 900-degree centigrade. It also should be included in the safety-blocking function.

6.  The radiation source is a porous ceramic heat radiating plate of 300mm*450mm and tilted 30 °.

7.  The pilot burner should be stationed equally as 0.7mm diameter and 19/EA produced a slop of 60 degrees per centerline bottom.

8.  Air velocity of exhaust stack should be attached an exhauster as 2.5±0.2 m/s.

9.  It should be used a radiation pyrometer which can be monitor 250mm diameter from about 1.4 meters and has ±5 degree centigrade sod ml accessory with 480℃ to 530℃ (black body temperature) range for adjustment the surface quality of heat of radiant panel.

10.  It can be measured hat flow within tolerance from radiant heat panel to 110mm to 910mm at 100mm intervals.

11.  It can be adjusting the flame length of 60mm to 120mm on Pilot burner flame.

12.  It can be adjust the limited temperature (900℃) of backside panel according to air flow condition of the radiation heat panel.

13.  Inlet flows control valve, precise control of propane flow.


Test Result:

•  Measure critical heat flux (CHF) which observed flame spread length transformers Kw/m2 using by heat streamline velocity. 

•  If specimen takes more than 30 minute testing time, it could record a flame extinction time and the farthest of flame spread, then transforms it’s into CHF. 

•  Need to smoke measurement section. it can be measured smoke density by the transmittance of the photometric system.



   Specification:   

1. Steel frame chamber, Dimension:1,400mm(W)×500mm(D)×2,200mm(H)

2. Console Size: 600mm(W)×750mm(D)×1600mm(H)

3. Power:AC 220V, 50/60Hz, 20A

4. Weight: about 300kg

5. Instructions: Provide

6. Exhaust:30~85m3/min

7. Calcium silicate insulation board

8. Observation window

9. Sliding platform for loading the specimen

10. Auto moving ignition burner

11. Scale for check flame spread

12. K-type thermocouples: Range, Ambient about 1500 degree centigrade

13. Stainless steel sample holder

14. Dummy calibration panel holder for measuring heat fluxmeter

15. Pyrometer for measuring radiant panel temperature

16. Utilities: Bottled propane gas, Air compressor (5HP)